๐Ÿ”ฌ EP26 Reagent Lot-to-Lot Verification Tool
CLSI EP26 Ed.2 ยท 2022 | Author: Dr. Raafat ElShimy
1 Protocol Setup โ€” CLSI EP26 Stage 1
Purpose: This tab establishes the protocol parameters for each analyte (one-time setup). Enter analyte data, then the tool calculates the Rejection Limit (RL) and required sample size (N) by looking up the embedded CLSI EP26 Appendix A tables (A1, A2, A3) based on CD/SWRL, Sr/SWRL, and desired statistical power. The output checklist is used for all future lot evaluations.
2 Step 1: Download Input Template

Download and complete the CSV template with your analyte parameters. Fill one row per analyte concentration level, then import it below.

3 Step 2: Enter or Import Analyte Data
Instructions:
  • CD: Critical Difference (absolute units or %TEa-based, same units as SWRL)
  • SWRL: Within-Reagent-Lot imprecision (total, SD)
  • Sr: Repeatability / within-run SD
  • Levels: Number of medical decision concentrations being tested (1, 2, or 3 โ†’ selects Table A1, A2, A3)
  • Power Goal: Desired statistical power (0.80, 0.85, 0.90, or 0.95)
# Analyte Conc. Level Label Target Conc. Units CD SWRL Sr No. of Levels Power Goal Action
๐Ÿ“Š

No analyte data entered yet. Click "Add Analyte Level" or import a CSV file.

4 Step 3: Calculate Protocol Parameters
Calculations use the embedded CLSI EP26 Appendix A tables. The tool computes CD/SWRL and Sr/SWRL ratios, selects the appropriate table (A1 for 1 level, A2 for 2, A3 for 3), and returns: Rejection Limit (RL) = k ร— CD, Number of Samples (N), and achieved Power.
๐Ÿงฎ

Click "Calculate All Parameters" after entering analyte data above.

1 Lot Evaluation โ€” CLSI EP26 Stage 2
Purpose: Enter measured results from the current and candidate reagent lots. The tool calculates the mean difference for each analyte concentration level and compares it against the Rejection Limit (RL). A pass/fail conclusion is generated for each level and analyte.
2 Step 1: Load Protocol Parameters

Load the checklist generated in Tab 1, or manually define analytes and their parameters for this evaluation.

๐Ÿ“‹

No protocol loaded. Load from Tab 1 or import a checklist CSV.

3 Step 2: Enter Measurement Results
Enter paired measurements for each sample: the result from the Current Lot and the result from the Candidate Lot. Difference = Candidate โˆ’ Current. Minimum samples per level as specified in the protocol (N required).
๐Ÿงช

Load protocol parameters first (Step 1 above).

4 Step 3: Evaluation Results & Conclusion
๐Ÿ“ˆ

Enter data and click "Calculate & Evaluate" to see results.